• Title of article

    Monolithic silica columns of various format in automated sample clean-up/multidimensional liquid chromatography/mass spectrometry for peptidomics

  • Author/Authors

    Machtejevas، نويسنده , , Egidijus and Andrecht، نويسنده , , Sven and Lubda، نويسنده , , Dieter and Unger، نويسنده , , Klaus K.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    5
  • From page
    97
  • To page
    101
  • Abstract
    The following particulate and monolithic silica columns were implemented in a fully automated and flexible multidimensional LC/MS system with integrated sample clean-up, to perform the analysis of endogeneous peptides from filtered urine and plasma samples: restricted access sulphonic acid strong cation-exchanger (RAM-SCX) for sample clean-up, RP 18 Chromolith guard columns as trap columns and 100 μm I.D. monolithic RP 18 fused silica capillary columns as last LC dimension. The results show sufficient overall system reproducibility and repeatability. Implementation of monolithic silica columns added an additional flexibility with respect to flow rate variation and adjustment due to the low column back pressures. Also, monolithic columns showed a lower clogging rate in long-term usage for biological samples as compared to particulate columns. The applied system set-up was tested to be useful for the routine peptide screening in search of disease biomarkers.
  • Keywords
    Peptide separation , Monolithic silica column
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2007
  • Journal title
    Journal of Chromatography A
  • Record number

    1523062