Title of article :
Choosing an equivalent replacement column for a reversed-phase liquid chromatographic assay procedure
Author/Authors :
Dolan، نويسنده , , J.W. and Maule، نويسنده , , A. and Bingley، نويسنده , , D. and Wrisley، نويسنده , , L. and Chan، نويسنده , , C.C. and Angod، نويسنده , , M. and Lunte، نويسنده , , C. and Krisko، نويسنده , , R. and Winston، نويسنده , , J.M. and Homeier، نويسنده , , B.A. and McCalley، نويسنده , , D.V. and Snyder، نويسنده , , L.R.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
16
From page :
59
To page :
74
Abstract :
The column selectivity parameters (H, S*, A, B and C) described in the preceding paper [L.R. Snyder, A. Maule, A. Heebsch, R. Cuellar, S. Paulson, J. Carrano, L. Wrisley C.C. Chan, N. Pearson, J.W. Dolan, J.J. Gilroy, J. Chromatogr. A 1057 (2004) 49–57] can be used to compare columns in terms of selectivity. A detailed procedure for such column comparisons is presented here, and evaluated by its use in finding suitable replacement columns for 12 different routine separations performed in five different pharmaceutical analysis laboratories.
Keywords :
Selectivity parameters , Steric resistance , Hydrogen-bond acidity , Column selectivity , Cation-exchange capacity , Hydrogen-bond basicity , Hydrophobicity
Journal title :
Journal of Chromatography A
Serial Year :
2004
Journal title :
Journal of Chromatography A
Record number :
1523627