Title of article
Computerized separation of chromatographically unresolved peaks
Author/Authors
Krup??k، نويسنده , , J. and Mydlov?، نويسنده , , J. and ?p?nik، نويسنده , , I. and Tienpont، نويسنده , , B. and Sandra، نويسنده , , P.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
10
From page
80
To page
89
Abstract
A computerized peak deconvolution software and mass spectra were successfully applied for the deconvolution of overlapped peak cluster in the chromatogram obtained separating the complex mixture of pesticides by retention time locking gas chromatography–mass spectroscopy. The method based on the unique fragment ions in the spectra can be used for deconvolution of peak clusters if mass spectra of overlapped peaks differ. This method allows determining actual retention times of overlapped peaks. Peak areas found by this method however, cannot be used naturally for the quantitative purposes as the abundance of fragment ions used for this deconvolution procedure can dramatically differ. Computer assisted deconvolution of peaks in the peak clusters gives more realistic peak area ratios as at this method it is supposed equal response for all peaks overlapped in a cluster.
Keywords
pesticides , Retention time locking (RTL) , Capillary GC–MS , Deconvolution of overlapped peaks , multi-residue analysis
Journal title
Journal of Chromatography A
Serial Year
2005
Journal title
Journal of Chromatography A
Record number
1524216
Link To Document