Title of article :
Sequential injection extraction based on restricted access material for determination of furosemide in serum
Author/Authors :
Huclov?، نويسنده , , Jitka and ?at?nsk?، نويسنده , , Dalibor and Maia، نويسنده , , Tiago and Karl??ek، نويسنده , , Rolf and Solich، نويسنده , , Petr and Ara?jo، نويسنده , , Alberto Nova، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
7
From page :
245
To page :
251
Abstract :
Restricted access material (RAM) column containing 25 μm C18 alkyl-diol support was integrated into the sequential injection analysis (SIA) manifold and the SIA–RAM system was tested for direct determination of furosemide in serum. LiChrospher® ADS column based on restricted access material is proposed to direct injection of biofluids. The integration of RAM material into SIA enabled creation of a comprehensive on-line sample clean-up technique combined with fluorescence quantitation of analyte. fuged and diluted serum sample was aspirated into the system and loaded onto the column using acetonitrile–water (2:98), pH 2.7. The analyte was retained on the column while proteins contained in the sample were removed to the waste without precipitation and clogging the column. Interfering substances complicating the detection were washed out by acetonitrile–water (15:85), pH 2.7 in the next step. The extracted analyte was eluted by means of acetonitrile–water (25:75), pH 2.3 to the fluorescence detector (emission filter 385 nm). The whole procedure comprising sample pre-treatment, analyte detection and column reconditioning took 20 min. The recoveries of furosemide from serum lay between 101.4 and 103.4% for three concentrations of analyte.
Keywords :
Furosemide , Restricted access materials , serum , Sequential injection analysis
Journal title :
Journal of Chromatography A
Serial Year :
2005
Journal title :
Journal of Chromatography A
Record number :
1524292
Link To Document :
بازگشت