• Title of article

    High-throughput analysis of tetracycline and penicillin antibiotics in animal tissues using electrospray tandem mass spectrometry with selected reaction monitoring transition

  • Author/Authors

    Goto، نويسنده , , Tomomi and Ito، نويسنده , , Yuko and Yamada، نويسنده , , Sadaji and Matsumoto، نويسنده , , Hiroshi and Oka، نويسنده , , Hisao، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    7
  • From page
    193
  • To page
    199
  • Abstract
    A simple, rapid, and simultaneous analysis method for oxytertracycline, tetracycline, chlortetracycline, penicillin G, ampicillin, and nafcillin in meat has been developed by using electrospray ionization tandem mass spectrometry. The sample preparation was performed by homogenizing with water followed by a centrifugal ultrafiltration, after addition of internal standards (demeclocycline, penicillin G-d5, ampicillin-d5 and nafcillin-d6). The MS/MS analysis involves the combined use of sample enrichment on the short column and a multiple reaction monitoring technique. The overall recoveries from animal (bovine and swine) muscle, kidney, and liver fortified at the levels of 0.05 and 0.1 ppm ranged from 70 to 115% with the coefficients of variation ranging from 0.7 to 14.8% (n = 5). Analysis time, including sample preparation and determination, is only 3 h per eight sample and detection limits for all antibiotics are 0.002 ppm. The method is considered to be satisfactory for the rapid screening of the tetracycline and penicillin antibiotic residues in meat.
  • Keywords
    ESI LC–MS/MS , Animal tissues , Tetracycline , Penicillins
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2005
  • Journal title
    Journal of Chromatography A
  • Record number

    1524586