Title of article :
Simultaneous determination of palladium, platinum, rhodium and gold by on-line solid phase extraction and high performance liquid chromatography with 5-(2-hydroxy-5-nitrophenylazo)thiorhodanine as pre-column derivatization regents
Author/Authors :
Hu، نويسنده , , Qiufen and Yang، نويسنده , , Xiangjun and Huang، نويسنده , , Zhangjie and Chen، نويسنده , , Jing and Yang، نويسنده , , Guangyu، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
6
From page :
77
To page :
82
Abstract :
In this paper, 5-(2-hydroxy-5-nitrophenylazo)thiorhodanine (HNATR) was synthesized. A new method for the simultaneous determination of palladium, platinum, rhodium and gold ions as metal-HNATR chelates was developed using a rapid analysis column high performance liquid chromatography equipped with on-line solid phase extraction technique. The samples (Water, human urine, geological samples and soil) were digested by microwave acid-digestion. The palladium, platinum, rhodium and gold ions in the digested samples were pre-column derivatized with HNATR to form colored chelates. The Pd-HNATR, Pt-HNATR, Rh-HNATR and Au-HNATR chelates can be absorbed onto the front of the enrichment column when they were injected into the injector and sent to the enrichment column [Zorbax Stable Bound, 10 mm × 4.6 mm, 1.8 μm] with a buffer solution of 0.05 mol L−1 phosphoric acid as mobile phase. After the enrichment had finished, by switching the six ports switching valve, the retained chelates were back-flushed by mobile phase and travelling towards the analytical column. These chelates separation on the analytical column [Zorbax Stable Bound, 10 mm × 4.6 mm, 1.8 μm] was satisfactory with 72% acetonitrile (containing 0.05 mol L−1 of phosphoric acid and 0.1% of Triton X-100) as mobile phase. The palladium, platinum, rhodium and gold chelates were separated completely within 2.5 min. Compared to the routine chromatographic method, more then 80% of separation time was shortened. By on-line solid phase extraction system, a large volume of sample (10 mL) can be injected, and the sensitivity of the method was greatly improved. The detection limits (S/N = 3, the sample injection volume is 10 mL) of palladium, platinum, rhodium and gold in the original samples reaches 1.4, 1.8, 2.0 and 1.2 ng L−1, respectively. The relative standard deviations for five replicate samples were 2.4–3.6%. The standard recoveries were 88–95%. This method was applied to the determination of palladium, platinum, rhodium and gold in human urine, water and geological samples with good results.
Keywords :
Rhodium , 5-(2-Hydroxy-5-nitrophenylazo)thiorhodanine , Rapid column high performance liquid chromatography , Gold , On-line solid phase extraction , PALLADIUM , Platinum
Journal title :
Journal of Chromatography A
Serial Year :
2005
Journal title :
Journal of Chromatography A
Record number :
1524617
Link To Document :
بازگشت