Title of article :
Reply to “Suitability of N,O-bis(trimethylsilyl)trifluoroacetamide as derivatization reagent for the determination of the estrogens estrone and 17α-ethinylestradiol by gas chromatography-mass spectrometry”
Author/Authors :
Shareef، نويسنده , , Ali and Angove، نويسنده , , Michael J. and Wells، نويسنده , , John D. and Johnson، نويسنده , , Bruce B.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
The HPLC retention times of several substituted phenols have been measured and simulated using Advanced Chemistry Developmentʹs LC simulator, using 50% acetonitrile (ACN) as the mobile phase. For alkyl- and nitro-substituted phenols, the quality of the simulation improves when pH of the mobile phase is estimated and used in the simulation. Simply using the pH of the buffer gives simulation results that are not as close to the actual retention times. However, the opposite is the case for halogenated phenols. The pKa values in 50% ACN for some of these phenols have also been determined, which tend to be one unit higher than the aqueous pKa values reported in the literature.
Keywords :
17?-Ethinylestradiol , Derivatization , O-bis(trimethyl)trifluoroacetamide , Estrone , N , Silylation , GC
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A