• Title of article

    Development of a removable multi-column-switching system for gas chromatography

  • Author/Authors

    Hinz، نويسنده , , Dirk-Christian Elsner، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    8
  • From page
    87
  • To page
    94
  • Abstract
    A removable column-switching device for a commercially available gas chromatograph has been developed. To this end, the challenges of the equipmentʹs high temperature stability, of the flow of gas through stand-by columns, of catalytically effective stainless steel surfaces, of the cooling trap effects within valves and the limited oven space are dealt with. The developed column-switching system was investigated in terms of dead volume (tailing factor), adsorption (Grob test) and degradation (Endrin test). This GC column-switching system shows comparable results with analyses that do not employ column switching. For example, it was possible to avoid effects, such as adsorption and decomposition. This system can automatically test the suitability of as many as eight separating columns to analyze unknown samples. The GC column-switching system thus serves to increase efficiency in the selection or screening of stationary phases with GC method developments.
  • Keywords
    Multiple column-switching device , Screening of stationary phases , Increased sample throughput , Method development , Automation , Capillary gas chromatography
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2006
  • Journal title
    Journal of Chromatography A
  • Record number

    1525124