Title of article :
Optimising reversed-phase liquid chromatographic separation of an acidic mixture on a monolithic stationary phase with the aid of response surface methodology and experimental design
Author/Authors :
Wang، نويسنده , , Y. and Harrison، نويسنده , , M. and Clark، نويسنده , , B.J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
An optimization strategy for the separation of an acidic mixture by employing a monolithic stationary phase is presented, with the aid of experimental design and response surface methodology (RSM).
hogonal array design (OAD) OA16 (215) was used to choose the significant parameters for the optimization. The significant factors were optimized by using a central composite design (CCD) and the quadratic models between the dependent and the independent parameters were built. The mathematical models were tested on a number of simulated data set and had a coefficient of R2 > 0.97 (n = 16). On applying the optimization strategy, the factor effects were visualized as three-dimensional (3D) response surfaces and contour plots. The optimal condition was achieved in less than 40 min by using the monolithic packing with the mobile phase of methanol/20 mM phosphate buffer pH 2.7 (25.5/74.5, v/v). The method showed good agreement between the experimental data and predictive value throughout the studied parameter space and were suitable for optimization studies on the monolithic stationary phase for acidic compounds.
Keywords :
Orthogonal array design , Evaluation of column , Response surface fitting , Central composite design , Chemometrics