Title of article :
Rapid chloride analysis using miniaturised isotachophoresis
Author/Authors :
Prest، نويسنده , , Jeff E. and Baldock، نويسنده , , Sara J. and Fielden، نويسنده , , Peter R. and Goddard، نويسنده , , Nicholas J. and Mohr، نويسنده , , Stephan and Treves Brown، نويسنده , , Bernard J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
5
From page :
183
To page :
187
Abstract :
A new design of miniaturised separation device for performing isotachophoresis (ITP) has been produced. The device contains a simple arrangement of channels comprising a single separation channel with a ‘double T’ injection geometry. The device was produced in poly(methyl methacrylate) and incorporates an on-column conductivity detector. A new electrolyte system was developed to enable the rapid determination of chloride to be made. This electrolyte system uses a leading ion of 3.5 mM nitrate at pH 3.0 with 0.5 mM indium(III) added as a complexing agent. Use of this electrolyte system with the new separation device allowed chloride samples to be analysed in under 100 s, with a limit of detection (LOD) calculated to be 2.2 mg l−1.
Keywords :
microdevice , CHIP , Chloride analysis , Conductivity detection , Isotachophoresis , Miniaturisation
Journal title :
Journal of Chromatography A
Serial Year :
2006
Journal title :
Journal of Chromatography A
Record number :
1525474
Link To Document :
بازگشت