Title of article :
Automated capillary gas chromatographic system to monitor ethylene emitted from biological materials
Author/Authors :
Pham-Tuan، نويسنده , , Hai and Vercammen، نويسنده , , Joeri and Devos، نويسنده , , Christophe and Sandra، نويسنده , , Pat، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
An automated capillary gas chromatographic system to measure ethylene emitted from biological materials is presented. The system consists of an on-line sampling device, a thermodesorption preconcentration apparatus and a capillary gas chromatograph with a flame ionization detection system. The limit of detection achievable on the GC system alone is 5 pg ethylene. The use of the strong Carboxen 1000 adsorbent at a sampling temperature as low as −50°C allows sampling of volumes up to a few liters. Ethylene concentrations at low ppt levels can be accurately and reproducibly determined.
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A