• Title of article

    Robustness testing of a reversed-phase high-performance liquid chromatographic assay: comparison of fractional and asymmetrical factorial designs

  • Author/Authors

    Hund، نويسنده , , E. and Vander Heyden، نويسنده , , Y. and Haustein، نويسنده , , M. and Massart، نويسنده , , D.L. and Smeyers-Verbeke، نويسنده , , J.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    19
  • From page
    167
  • To page
    185
  • Abstract
    Robustness tests were performed on a reversed-phase HPLC assay for triadimenol. Different experimental designs were compared. Two-level fractional factorial designs with different resolutions were used to study the influence of procedure-related factors. The factors chromatographic column manufacturer at four levels and instrument at three levels were stepwise included in the study using asymmetrical factorial designs. The significance of the factor effects was determined statistically, using two types of error estimates in the calculation of critical effects, and graphically, by means of half-normal plots. The asymmetrical designs turned out to be an efficient and economic method to examine the influence of factors at different numbers of levels in the robustness testing of analytical methods.
  • Keywords
    Half-normal plots , Critical effects , Robustness test , Fractional factorial design
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2000
  • Journal title
    Journal of Chromatography A
  • Record number