Title of article :
Determination of phenolic antioxidants in aviation jet fuel
Author/Authors :
Bernabei، نويسنده , , M and Bocchinfuso، نويسنده , , G and Carrozzo، نويسنده , , P and De Angelis، نويسنده , , C، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
The world-wide aviation jet fuel used for civil and military aircraft is of a kerosene type. To avoid peroxide production after the refinery process a specific antioxidant additive should be added on fuel. The antioxidants generally used are based on hindered phenols in a range of concentration 10–20 μg/ml. In the present work a specific method to measure the concentration of phenolic antioxidants is shown. The method is based on a liquid chromatographic technique with electrochemical detection. The technique, because of its selectivity, does not require sample pre-treatments. The analysis of a 5–10 ml fuel sample can be performed in less than 10 min with a sensitivity of 0.1 μg/ml and a RSD=2.5%. A comparison with another highly selective gas chromatographic technique with mass spectrometric detection with selected ion monitoring (GC–MS-SIM) is reported. The sensitivity of GC–MS-SIM method was 2 μg/ml with a RSD=3.1%.
Keywords :
antioxidants , Phenols , Alkylphenols
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A