Title of article :
Monte Carlo calculations of AAPM Task Group Report No. 43 dosimetry parameters for the MED3631-A/M 125I source
Author/Authors :
Rivard، Mark J. نويسنده ,
Issue Information :
فصلنامه با شماره پیاپی سال 2015
Abstract :
Brachytherapy dosimetry parameters for MED3631-A/M 125I sources have been determined in accordance with the AAPM Task Group No. 43 (TG-43) dosimetry protocol. These data were calculated using the*F8 tally from the MCNP4B2 Monte Carlo radiation transport code with the DLC-189 cross-section libraries. Due to motion of the 125I resin beads and gold-copper markers within the capsule, parameters such as the geometry function, radial dose function, dose rate constant, and anisotropy function were examined with the beads and markers having either "realistic" or "ideal" positions; the realistic position was a weighted combination of "vertical" and "diagonal" capsule orientations.
Keywords :
short circuit current , transient over voltage , Fault current limiter , power quality