Title of article :
Wavelength dependence of the precision of noninvasive optical measurement of oxy-, deoxy-, and total-hemoglobin concentration
Author/Authors :
Yamashita، Yuichi نويسنده , , Maki، Atsushi نويسنده , , Koizumi، Hideaki نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
-1107
From page :
1108
To page :
0
Abstract :
The precision of noninvasive optical measurement of the concentration changes in oxy-, deoxy-, and total-hemoglobin depends on wavelength. For estimating the precision, we calculated the noise level of the concentration changes as the uncertainty in measurements using several wavelength pairs of light. Seven laser diodes (664-848 nm) were used simultaneously for spectroscopic measurement of brain activity during finger motor stimulation. We also used the analysis of error propagation from the uncertainty in direct measurements of absorbance changes to estimate indirectly the uncertainty of concentration changes. The measurement of the concentration changes made using an 830/664-nm pair are two times (oxy-Hb) and six times (deoxy-Hb) more precise than those made using an 830/782-nm pair.