Title of article :
Ridge filter design and optimization for the broad-beam three-dimensional irradiation system for heavy-ion radiotherapy
Author/Authors :
Schaffner، Barbara نويسنده , , Kanai، Tatsuaki نويسنده , , Futami، Yasuyuki نويسنده , , Shimbo، Munefumi نويسنده , , Urakabe، Eriko نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2017
Abstract :
A method for designing a variable-SOBP (spread-out Bragg peak) ridge filter has been proposed. First, ridge filter parameters are determined by using a Monte Carlo calculation followed by a fast two-step iterative optimization. Then, tilting the ridge filter results in continuous variation of the SOBP width. Monte Carlo calculations show that depth dose uniformity changes from ±1.3% to ± 1.6% for SOBP widths ranging from 10.3 cm to 14.5 cm. Advantages of the proposed tilting ridge filter include a capability of continuous SOBP variation and cost-effective installation for a given SOBP width range.
Keywords :
multiple field optimization , heavy-ion radiotherapy , ridge filter , intensity modulation
Journal title :
MEDICAL PHYSICS