Title of article :
Intensity modulated radiotherapy with charged particle beams: Studies of inverse treatment planning for rotation therapy
Author/Authors :
Bortfeld، T. نويسنده , , Oelfke، U. نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2015
Abstract :
A method for designing a variable-SOBP (spread-out Bragg peak) ridge filter has been proposed. First, ridge filter parameters are determined by using a Monte Carlo calculation followed by a fast two-step iterative optimization. Then, tilting the ridge filter results in continuous variation of the SOBP width. Monte Carlo calculations show that depth dose uniformity changes from ±1.3% to ± 1.6% for SOBP widths ranging from 10.3 cm to 14.5 cm. Advantages of the proposed tilting ridge filter include a capability of continuous SOBP variation and cost-effective installation for a given SOBP width range.
Keywords :
radiation therapy , charged particle beams , inverse planning , rotation therapy , intensity modulation , distal edge tracking
Journal title :
MEDICAL PHYSICS
Journal title :
MEDICAL PHYSICS