Author/Authors :
Acebal، نويسنده , , Carolina C. and Insausti، نويسنده , , Matيas and Pistonesi، نويسنده , , Marcelo F. and Lista، نويسنده , , Adriana G. and Band، نويسنده , , Beatriz S. Fernلndez، نويسنده ,
Abstract :
The advantages of the flow-batch methodology were exploited to implement a simple system with nephelometric detection for the determination of monosodium glutamate (MSG) in food samples. The method is based on the inhibitory effect of the MSG over the crystallization of l-lysine in an isopropanol/acetone mixture. The calibration curve was prepared on-line. The method was linear over the range of 2.8 × 10−3 to 1.1 × 10−2 g L−1 and a detection limit of 9.7 × 10−5 g L−1 was achieved. It was successfully applied to determine the MSG concentration in food samples, without a previous treatment. A recovery study was carried out on real samples and the percentages were between 98 and 106%.