• Title of article

    Determination of selenium species in human urine by high performance liquid chromatography and inductively coupled plasma mass spectrometry

  • Author/Authors

    Quijano، نويسنده , , M.Angeles and Gutiérrez، نويسنده , , Ana Maria and Pérez-Conde، نويسنده , , M.Concepciَn and Cلmara، نويسنده , , Carmen، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 1999
  • Pages
    9
  • From page
    165
  • To page
    173
  • Abstract
    A method developed to determine organic and inorganic selenium species in human urine samples is presented in detail. After a simple sample treatment based on elimination of non-charged organic compounds, selenium species were separated by high performance liquid chromatography (HPLC) on a Spherisorb 5 ODS/AMINO column using two different chromatographic conditions: phosphate buffers at pH 2.8 and 6.0. Detection was carried out using an on-line inductively coupled plasma mass spectrometer (ICP-MS). Trimethylselenonium ion and two unknown selenium species in urine samples were found. Selenium species were shown to have stability problems, with the maximum allowed storage time of 1 week.
  • Keywords
    inductively coupled plasma mass spectrometry , Human urine , High Performance Liquid Chromatography , Selenium speciation
  • Journal title
    Talanta
  • Serial Year
    1999
  • Journal title
    Talanta
  • Record number