Title of article
Trace metals analysis in estuarine and seawater by ICP-MS using on line preconcentration and matrix elimination with chelating resin
Author/Authors
Nicola??، نويسنده , , M. and Rosin، نويسنده , , C. and Tousset، نويسنده , , N. and Nicolai، نويسنده , , Y.، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 1999
Pages
12
From page
433
To page
444
Abstract
The main difficulties of trace metals analysis in estuarine and seawater stem from their very low concentration (μg/l to sub-μg/l), and, by contrast, the high salt content (up to 38 g/l in the Mediterranean Sea). ICP-MS allows multi-elemental analysis and offers great sensitivity, but may be strongly affected by matrix effects induced by high salt contents (> 1 g/l). To perform trace metals analysis both in riverine, estuarine and seawater, we have developed a hyphenated method: ion chelation chromatography coupled on-line with ICP-MS. Iminodiacetate resin, Metpac CC-1 (Dionex), was used to concentrate most of the trace metals, and to separate them from alkaline and alkaline-earth metals. Behaviour of 17 elements (Pb, Cu, Cd, Ni, U, Cr, Mn, Al, Co, Ga, In, Zn, V, Tl, Bi, Ag and Sn) towards the resin was qualitatively investigated. A method validation, partly derived from AFNOR standard XPT 90-210, was carried out on 12 elements (Pb, Cu, Cd, Ni, U, Cr, Mn, Al, Co, Ga, Bi and In). Replicate measurements of multi-elemental standard solutions were used to check linearity, and to determine repeatability and detection limits. Method accuracy was then assessed by analysing two certified materials: a synthetic freshwater (SRM 1643d), and a natural filtered coastal seawater (NRCC CASS-3). An application assay of natural samples from the Rhône river (France) was eventually carried out, and the analytical results were found to be consistent with previous works.
Keywords
Seawater , Chelating resin , Metals , ICP-MS
Journal title
Talanta
Serial Year
1999
Journal title
Talanta
Record number
1639426
Link To Document