Title of article
An alternative method for the measurement of mineral sulphide in wastewater
Author/Authors
Pouly، نويسنده , , F. and Touraud، نويسنده , , E. and Buisson، نويسنده , , J.-F. and Thomas، نويسنده , , O.، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 1999
Pages
6
From page
737
To page
742
Abstract
A simple and rapid procedure for sulphide measurement in crude oil refinery wastewater has been developed. This method is based on the knowledge of the UV response of sulphur compounds and the mathematical deconvolution of the sample spectrum using reference spectra (specific compounds or aggregate spectra). The detection limit of the method is 0.5 mg l−1 for a quartz cell pathlength of 10 mm. The range is up to 15 mg l−1. The method has been validated for crude oil refinery wastewater. Compared to conventional methods, the UV determination is quicker and easier to run.
Keywords
Total sulphide , UV spectrophotometry , Industrial wastewater
Journal title
Talanta
Serial Year
1999
Journal title
Talanta
Record number
1639524