• Title of article

    Direct determination of cadmium and copper in seawater using a transversely heated graphite furnace atomic absorption spectrometer with Zeeman-effect background corrector

  • Author/Authors

    Chan، نويسنده , , Mei-Shu and Huang، نويسنده , , Shang-Da، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2000
  • Pages
    8
  • From page
    373
  • To page
    380
  • Abstract
    Methods for the direct determination of copper and cadmium in seawater were described using a graphite furnace atomic absorption spectrometer (GFAAS) equipped with a transversely heated graphite atomizer (THGA) and a longitudinal Zeeman effect background corrector. Ammonium nitrate was used as the chemical modifier to determine copper. The mixture of di-ammonium hydrogen phosphate and ammonium nitrate was used as the chemical modifier to determine cadmium. The matrix interference was removed completely so that a simple calibration curve method could be applied. This work is the first one with the capability of determining cadmium in unpolluted seawater directly with GFAAS using calibration curve based on simple aqueous standards. The accuracy of the methods was confirmed by analysis of three kinds of certified reference saline waters. The detection limits (LODs), with injection of a 20-μl aliquot of seawater sample, were 0.06 μg l−1 for copper and 0.005 μg l−1 for cadmium.
  • Keywords
    Seawater , Cadmium , Copper , ETAAS , Direct determination , GFAAS
  • Journal title
    Talanta
  • Serial Year
    2000
  • Journal title
    Talanta
  • Record number