Title of article :
Optimising mobile phase composition, its flow-rate and column temperature in HPLC using taboo search
Author/Authors :
Guillaume، نويسنده , , Yves Claude and Peyrin، نويسنده , , Eric، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2000
Pages :
8
From page :
579
To page :
586
Abstract :
A chemometric methodology is proposed to study the separation of seven p-hydroxybenzoic esters in reversed phase liquid chromatography (RPLC). Fifteen experiments were found to be necessary to find a mathematical model which linked a novel chromatographic response function (CRF) with the column temperature, the water fraction in the mobile phase and its flow rate. The CRF optimum was determined using a new algorithm based on Glover’s taboo search (TS). A flow-rate of 0.9 ml min−1 with a water fraction of 0.64 in the ACN–water mixture and a column temperature of 10°C gave the most efficient separation conditions. The usefulness of TS was compared with the pure random search (PRS) and simplex search (SS). As demonstrated by calculations, the algorithm avoids entrapment in local minima and continues the search to give a near-optimal final solution. Unlike other methods of global optimisation, this procedure is generally applicable, easy to implement, derivative free, conceptually simple and could be used in the future for much more complex optimisation problems.
Keywords :
Optimisation , chromatography , Taboo search , Chemometric