Title of article :
Highly sensitive and selective measurements of cobalt by catalytic adsorptive cathodic stripping voltammetry
Author/Authors :
Safavi، نويسنده , , A and Shams، نويسنده , , E، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2000
Abstract :
A very sensitive and selective catalytic adsorptive cathodic stripping procedure for trace measurements of cobalt is presented. The method is based on adsorptive accumulation of the cobalt-MTB (methyl thymol blue) complex onto a hanging mercury drop electrode, followed by reduction of the adsorbed species by voltammetric scan using differential pulse modulation. The reduction current is enhanced catalytically by nitrite. The optimum conditions for the analysis of cobalt include pH 9.0 (ammonia buffer), 2.0 μM methyl thymol blue, 0.8 M sodium nitrite and an accumulation potential of −0.5 V (versus Ag/AgCl). The peak current is proportional to the concentration of cobalt over the entire concentration range tested (0.02–500 ng ml−1) with a detection limit of 0.005 ng ml−1 for an accumulation time of 60 s. The method was applied to determination of cobalt in a mineral water sample and some analytical grade salts with satisfactory results.
Keywords :
Catalytic adsorptive stripping voltammetry , Cobalt , Methyl thymol blue