• Title of article

    Evaluation of different SIA sample–buffer configurations using a fluoride-selective membrane electrode as detector

  • Author/Authors

    van Staden، نويسنده , , Jacobus (Koos) F and Stefan، نويسنده , , Raluca-Ioana and Birghila، نويسنده , , Semaghiul، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2000
  • Pages
    9
  • From page
    3
  • To page
    11
  • Abstract
    The peak profiles of four different buffer–sample SIA configurations, e.g. buffer–sample; sample–buffer; buffer–sample–buffer and sample–buffer–sample with the last two in the sandwich mode were evaluated with a fluoride-selective membrane electrode as detector. The best response characteristics and peak shapes as well as recovery and precision values were obtained for the buffer–sample configuration. For low concentration levels, sandwich SIA configurations are more suitable, when optimum buffer and pH are used. The utilisation of a cheap electrolyte with a minimum consumption of buffer and a cheap, robust instrumentation made the SIA system suitable for on-line determination of fluoride.
  • Journal title
    Talanta
  • Serial Year
    2000
  • Journal title
    Talanta
  • Record number

    1640217