• Title of article

    Polypyrrole–calcion film as a membrane and solid-contact in an indicator electrode for potentiometric titrations

  • Author/Authors

    Blaz، نويسنده , , T and Migdalski، نويسنده , , J and Lewenstam، نويسنده , , A، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2000
  • Pages
    10
  • From page
    319
  • To page
    328
  • Abstract
    This paper shows the application of conducting polymers (CPs) for constructing potentiometric indicator electrodes. Two types of polypyrrole (PPy)-based calcium sensors are presented, one sensor with PPy–calcion film as the active part and the other sensor with PPy–calcion as a solid-state contact coated with a conventional membrane selective towards calcium ions. It is shown that the PPy–calcion film, due to the complexing properties of calcion ensuring high loading of the film with calcium, is sufficiently selective to be used as the active part or as a mediating layer of the indicator electrode. The electrode, with PPy–calcion film as the active part, was used as the indicator electrode in potentiometric titrations of calcium in mixed solvents, where conventional PVC-based electrode can not be used. For the first time, the practical applicability of PPy-based electrodes in titrations is demonstrated.
  • Keywords
    Titration , Polypyrrole–calcion film , Indicator Electrode , Potentiometric sensor
  • Journal title
    Talanta
  • Serial Year
    2000
  • Journal title
    Talanta
  • Record number

    1640337