Title of article :
Preconcentration and flame atomic absorption spectrometry determination of cadmium in mussels by an on-line continuous precipitation–dissolution flow system
Author/Authors :
Yebra، نويسنده , , M.C and Enr??quez، نويسنده , , M.F and Cesp?n، نويسنده , , R.M، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2000
Pages :
6
From page :
631
To page :
636
Abstract :
A rapid, sensitive, accurate and precise flame atomic absorption method is described for the determination of cadmium in mussels. The method is based on the continuous precipitation of cadmium as an ion pair between tetraiodocadmate and quinine and dissolution of the precipitate with ethanol. The metal can be preconcentrated 32-fold using 15 ml of sample solution by using a time-based technique at a sampling flow rate of 3.0 ml min−1. The proposed method allows the determination of cadmium in the range 0.25–5.5 μg g−1. The precision (relative standard deviation) obtained for different amounts of cadmium is in the range 1.5–4.7% at the 0.25–5.0 μg g−1 level. The method demonstrates high tolerance to interferences, and the data obtained are in agreement with the certified value of a selected reference material. This procedure was applied to the determination of cadmium in mussel samples from estuaries in Galicia (Spain).
Keywords :
Flame atomic absorption spectrometry , Precipitation and dissolution , Cadmium , Mussels , Flow Analysis
Journal title :
Talanta
Serial Year :
2000
Journal title :
Talanta
Record number :
1640461
Link To Document :
بازگشت