Title of article :
Construction and evaluation of ion selective electrodes for nitrate with a summing operational amplifier. Application to tobacco analysis
Author/Authors :
Pérez-Olmos، نويسنده , , R and Rios، نويسنده , , A and Fernلndez، نويسنده , , J.R and Lapa، نويسنده , , R.A.S and Lima، نويسنده , , J.L.F.C. Lima، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2001
Abstract :
In this paper, the construction and evaluation of an electrode selective to nitrate with improved sensitivity, constructed like a conventional electrode (ISE) but using an operational amplifier to sum the potentials supplied by four membranes (ESOA) is described. The two types of electrodes, without an inner reference solution, were constructed using tetraoctylammonium bromide as sensor, dibutylphthalate as solvent mediator and PVC as plastic matrix, the membranes obtained directly applied onto a conductive epoxy resin support. After the comparative evaluation of their working characteristics they were used in the determination of nitrate in different types of tobacco. The limit of detection of the direct potentiometric method developed was found to be 0.18 g kg−1 and the precision and accuracy of the method, when applied to eight different samples of tobacco, expressed in terms of mean R.S.D. and average percentage of spike recovery was 0.6 and 100.3%, respectively. The comparison of variances showed, on all ocassions, that the results obtained by the ESOA were similar to those obtained by the conventional ISE, but with higher precision. Linear regression analysis showed good agreement (r=0.9994) between the results obtained by the developed potentiometric method and those of a spectrophotometric method based on brucine, adopted as reference method, when applied simultaneously to 32 samples of different types of tobacco.
Keywords :
Potentiometry , Nitrate-selective electrode , Tobacco