Title of article
Redox-potentiometric determination of peroxide value in edible oils without titration
Author/Authors
Kardash-Strochkova، نويسنده , , E and Turʹyan، نويسنده , , Ya.I and Kuselman، نويسنده , , I، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2001
Pages
6
From page
411
To page
416
Abstract
Standard methods for peroxide value determination in edible oils, which are based on KI oxidation by the hydroperoxides and volumetric titration of the liberated iodine, have been improved using redox-potentiometric iodine determination without titration. Stages of the KI oxidation and redox-potentiometric measurements were combined in the same electrochemical cell. The limit of quantitation of the new method is 0.16 meq kg−1, which allows the analysis of fresh refined oils. The method is simple, not time and labor consuming, and suitable for automation.
Keywords
Peroxide value , Edible oils , metrological characteristics , redox-potentiometric method