• Title of article

    Flow injection on-line dilution for multi-element determination in human urine with detection by inductively coupled plasma mass spectrometry

  • Author/Authors

    Wang، نويسنده , , Jianhua and Harald Hansen، نويسنده , , Elo and Gammelgaard، نويسنده , , Bente، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2001
  • Pages
    10
  • From page
    117
  • To page
    126
  • Abstract
    A simple flow injection on-line dilution procedure with detection by inductively coupled plasma mass spectrometry (ICP-MS) was developed for the determination of copper, zinc, arsenic, lead, selenium, nickel and molybdenum in human urine. Matrix effects were minimized by employing a dilution factor of 16.5 with on-line standard addition, and 103Rh was used as internal standard to compensate for signal fluctuation. The procedure was validated by the analysis of two standard reference materials SRM 2670 (NIST) and Seronorm™ Trace Elements in Urine. Recovery experiments were performed by spiking the reference materials as well as artificial urine. The detection limits (μg l−1) were 0.12, 0.96, 0.30, 0.09, 0.45, 0.08, 0.09, and the precisions (RSD,%) were 2.6, 2.3, 3.0, 3.7, 3.7, 4.9, 2.8 for Cu, Zn, As, Pb, Se, Ni and Mo, respectively. The procedure was applied to the analysis of 41 human urine samples. No correlations between the concentrations of the elements were observed.
  • Keywords
    Flow Injection , ICP-MS , urine , On-line dilution
  • Journal title
    Talanta
  • Serial Year
    2001
  • Journal title
    Talanta
  • Record number