Title of article :
Determination of cadmium and lead in biological samples by Zeeman ETAAS using various chemical modifiers
Author/Authors :
Acar، نويسنده , , Orhan، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2001
Abstract :
The electrothermal atomic absorption spectrometric determination of cadmium and lead in biological certified reference materials (CRMs) has been carried out by using NH4H2PO4, Ni, Pd, Ni+NH4H2PO4, Pd+NH4H2PO4 and Ni+Pd+NH4H2PO4 as chemical modifiers. A comprehensive comparison was made among the modifiers in 1% Triton X-100 plus 0.2% nitric acid as diluent and without modifier. Zeeman background correction and graphite tubes inserted with platforms were used. Comparison was made in terms of pyrolysis and atomization temperatures, atomization and background absorption profiles. Ni+Pd+NH4H2PO4 modifier mixture was found to be preferable for the determination of Cd and Pb. Pyrolysis temperatures of analytes were increased up to 900 °C for Cd and 1250 °C for Pb by using Ni+Pd+NH4H2PO4 in 1% Triton X-100 plus 0.2% nitric acid diluent solution. Biological CRMs were analyzed to verify the accuracy and precision of this method. Depending on the biological sample type, the percent recoveries were increased from 62 to 102% for Cd and from 58 to 106% for Pb by using the proposed modifier mixture. The detection limits of Cd and Pb were found to be 0.04, 0.92 μg l−1, respectively.
Keywords :
Lead , Chemical modifiers , Cadmium , Zeeman background correction