Title of article :
On-line preconcentration and determination of copper, lead and chromium(VI) using unloaded polyurethane foam packed column by flame atomic absorption spectrometry in natural waters and biological samples
Author/Authors :
Anthemidis، نويسنده , , Aristidis N. and Zachariadis، نويسنده , , George A. and Stratis، نويسنده , , John A، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2002
Pages :
10
From page :
831
To page :
840
Abstract :
A simple, sensitive and low cost, flow injection time-based method was developed for on-line preconcentration and determination of copper, lead and chromium(VI) at sub μg l−1 levels in natural waters and biological samples. At the optimum pH, the on-line formed metal–ammonium pyrrolidine dithiocarbamate (APDC) complexes were sorbed on the unloaded commercial polyurethane foam (PUF), and subsequent eluted quantitatively by isobutylmethylketone and determined by flame atomic absorption spectrometry (FAAS). All chemical, and flow injection variables were optimized for the quantitative preconcentration of each metal and a study of interference level of various ions was also carried out. The system offered improved flexibility, low backpressure and applicability to all the studied metals. At a sample frequency of 36 h−1 and a 60 s preconcentration time, the enhancement factor was 170, 131 and 28, the detection limit was 0.2, 1.8 and 2.0 μg l−1, and the precision, expressed as relative standard deviation (sr), was 2.8 (at 10 μg l−1), 3.4 (at 50 μg l−1) and 3.6% (at 50 μg l−1) for Cu(II), Pb(II) and Cr(VI), respectively. The accuracy of the developed method was sufficient and evaluated by the analysis of certified reference materials and spiked water samples. Finally, the method was applied to the analysis of environmental samples.
Keywords :
Chromium , Atomic absorption spectrometry , On-line preconcentration , Copper , Polyurethane foam , Lead
Journal title :
Talanta
Serial Year :
2002
Journal title :
Talanta
Record number :
1644120
Link To Document :
بازگشت