• Title of article

    On-line separation, simultaneous dilution and spectrophotometric determination of zinc in fertilisers with a sequential injection system and xylenol orange as complexing agent

  • Author/Authors

    van Staden، نويسنده , , J.(Koos) F and Tlowana، نويسنده , , S.(Supi) I، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2002
  • Pages
    8
  • From page
    1115
  • To page
    1122
  • Abstract
    A dialyser unit, equipped with a passive neutral membrane, was incorporated into the conduits of a sequential injection (SI) system for the on-line removal of suspended solids and simultaneous dilution of the analyte before reaction and detection of the analyte. The system was applied to the determination of zinc(II) in fertilisers. The fully automated system is able to analyse zinc at a sampling frequency of ten samples per hour at a %R.S.D. of better than 0.55. The calibration graph was linear between 10 and 50 mg l−1. The detection limit was found to be 4.75 mg l−1. The results obtained with the proposed SI analyser compared favourably with the standard manual flame atomic absorption spectrometric method.
  • Keywords
    On-line dialysis , Sequential injection analysis , Zinc , fertilizers
  • Journal title
    Talanta
  • Serial Year
    2002
  • Journal title
    Talanta
  • Record number