Title of article :
Automated sequential injection fluorimetric determination and dissolution studies of Ergotamine Tartrate in pharmaceuticals
Author/Authors :
Legnerov?، نويسنده , , Sklenarova، Hana نويسنده , , H and Solich، نويسنده , , P، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2002
Pages :
5
From page :
1151
To page :
1155
Abstract :
A fully automated flow system for drug-dissolution studies based on the sequential injection analysis (SIA) was described and used for monitoring dissolution profiles of Ergotamine Tartrate (ET) in pharmaceutical formulations. 50 μl of dissolution medium was taken for each measurement at a flow rate of 40 μl s−1 and detected by fluorescence detector using λex=236 nm (λem≥390 nm). The calibration curve was linear over the range 0.03–0.61 mg l−1 of ET (sufficient for the dissolution tests). Equation of the calibration curve was calculated giving the following values: F=117.7 c+0.80 (n=6); r=0.9998. Detection limit was 0.01 mg l−1 of ET. The R.S.D. is less than 0.54 and 0.86% (n=10) when determining 0.61 and 0.03 mg l−1 of ET in standard solution, respectively. The dissolution test of Bellaspon tablets (0.3 mg of ET in 1 tablet) was programmed for 20 min, with a continuous sampling rate of 120 h−1 under conditions required by BP 1993.
Keywords :
Sequential injection analysis (SIA) , Ergotamine Tartrate , Dissolution test , Fluorimetric detection