Title of article :
Stopped-flow injection simultaneous determination of phosphate and silicate using molybdenum blue
Author/Authors :
Grudpan، نويسنده , , K. and Ampan، نويسنده , , P. and Udnan، نويسنده , , Y. and Jayasvati، نويسنده , , S. and Lapanantnoppakhun، نويسنده , , S. and Jakmunee، نويسنده , , J. and Christian، نويسنده , , G.D. and Ruzicka، نويسنده , , J.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2002
Pages :
8
From page :
1319
To page :
1326
Abstract :
Kinetic information for the phosphate–molybdate–ascorbic acid reaction can be obtained by making use of a very simple manually operated stopped-flow injection (FI) system. Various parameters (concentrations of reagents, flow rate, mixing coils, and volume of flow cell) were investigated for determination of phosphate. A stopped-FI system should be arranged for low degree of mixing (of reactants) and low dispersion so that good signals of rate changes will be observed. Simultaneous determination of phosphate and silicate by the stopped-FI technique is proposed, using a laboratory-made semi-automatic stopped-FI Analyzer with LED-based photometer. It is based on kinetic separation of phosphate and silicate using molybdenum blue. The proposed procedure has been demonstrated for the application to water samples. The results obtained agree with that of a standard method.
Keywords :
Stopped-FI , Simultaneous determination , silicate , Phosphate , Molybdenum blue , Kinetics
Journal title :
Talanta
Serial Year :
2002
Journal title :
Talanta
Record number :
1644254
Link To Document :
بازگشت