• Title of article

    Voltammetry determination of trace manganese with pretreatment glassy carbon electrode by linear sweep voltammetry

  • Author/Authors

    Di، نويسنده , , Junwei and Zhang، نويسنده , , Feng، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2003
  • Pages
    6
  • From page
    31
  • To page
    36
  • Abstract
    This paper described the determination of trace manganese using linear sweep voltammetry at a pretreatment glassy carbon electrode. The glassy carbon electrode pretreated by electrochemical method in the 0.1 mol l−1 NaOH solution greatly improved the electrode responsibility in the determination of manganese(II). The barrier to the detection of low manganese concentration was overcome by means of autocatalytic effect of manganese oxide deposited on the electrode in advance. Under the optimum experiments condition (0.04 mol l−1 NH3–NH4Cl buffer solution, pH 9.0), the linear range was 4×10−8 to 1×l0−6 mol l−1 Mn(II) for linear sweep voltammetry and 1×10−9 to 4×10−8 mol l−1 Mn(II) for convolution voltammetry. The relative standard deviation for 2×10−8 mol l−1 Mn(II) is 3.4%. The proposed method is simple, rapid, sensitive and selective. It had been applied to the determination of trace manganese in samples with satisfactory results.
  • Keywords
    Manganese , Linear sweep voltammetry , Electrochemical pretreatment , Autocatalytic effect , Glassy carbon electrode
  • Journal title
    Talanta
  • Serial Year
    2003
  • Journal title
    Talanta
  • Record number