• Title of article

    Real-time immunoassay of ferritin using surface plasmon resonance biosensor

  • Author/Authors

    Cui، نويسنده , , Xiaoqiang and Yang، نويسنده , , Fan and Sha، نويسنده , , Yufang and Yang، نويسنده , , Xiurong، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2003
  • Pages
    9
  • From page
    53
  • To page
    61
  • Abstract
    A surface plasmon resonance (SPR) biosensor was used for the first time to determine the concentration of ferritin in both HBS-EP buffer and serum. The monoclonal antibody was immobilized on the carboxymethyl dextran-modified gold surface by an amine coupling method. The interaction of antibody with antigen was monitored in real-time. The signal was enhanced by sandwich amplification strategy to improve the sensitivity and specificity of the immunoassay, especially in serum. The linear range of the assay in serum is over 30–200 ng ml−1 with the detection limit of 28 ng ml−1. The sensitivity, specificity, and reproducibility of the assay are satisfactory. The analyte and enhancement antibody-binding surface could be regenerated by pH 2.0 glycine–HCl buffer and the same antibody-immobilized surface could be used for more than 50 cycles of ferritin binding and regeneration.
  • Keywords
    Immunoassay , surface plasmon resonance , Sandwich enhancement , Ferritin
  • Journal title
    Talanta
  • Serial Year
    2003
  • Journal title
    Talanta
  • Record number