• Title of article

    Determination of l- and d-enantiomers of methotrexate using amperometric biosensors

  • Author/Authors

    Stefan، نويسنده , , Raluca-Ioana and Bokretsion، نويسنده , , Rahel Girmai and van Staden، نويسنده , , Jacobus Frederick and Aboul-Enein، نويسنده , , Hassan Y، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2003
  • Pages
    8
  • From page
    983
  • To page
    990
  • Abstract
    In order to determine the enantiopurity of methotrexate (Mtx), seven biosensors were proposed for the assay of l-Mtx and three biosensors for the assay of d-Mtx. The biosensors were designed using physical and chemical immobilization of glutamate oxidase and/or l-amino acid oxidase (l-AAOD) and/or horseradish peroxidase (HRP) for the assay of l-methotherexate, and d-amino acid oxidase (d-AAOD) and HRP for the assay of d-Mtx. Electrode characteristics were obtained and compared for the different carbon paste based biosensors. The linear concentration ranges for the proposed biosensors were in the ranges of fmol l−1 to pmol l−1, magnitude order with limits of detection in the fmol l−1 to nmol l−1 concentration range. All biosensors were successful for the determination of the enantiopurity of Mtx as raw material, and in its pharmaceutical formulations (tablets and injections).
  • Keywords
    l(d)-Mtx , Amperometric biosensor , Horseradish peroxidase , l(d)-Amino acid oxidase , Glutamate oxidase
  • Journal title
    Talanta
  • Serial Year
    2003
  • Journal title
    Talanta
  • Record number