Title of article :
Novel flow injection-fluorometric method for the determination of trace silicate and its application to ultrapurified water analysis
Author/Authors :
Sabarudin، نويسنده , , Akhmad and Oshima، نويسنده , , Mitsuko and Ishii، نويسنده , , Naoe and Motomizu، نويسنده , , Shoji، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2003
Abstract :
A highly sensitive fluorescence quenching method for the determination of silicate based on the formation of an ion associate between molybdosilicate and Rhodamine B (RB) in nitric acid medium was developed. A flow injection system coupled with a fluorescence detector was used for the measurement of fluorescence intensity at 560 and 580 nm as excitation and emission wavelengths, respectively. The calibration graph for Si showed a linear range of 0.1–5 ng cm−3 with correlation coefficient of 0.9999, and the detection limit of 0.06 ng cm−3. The proposed method was successfully applied to the determination of silicate in ultrapurified water with satisfactory results.
Keywords :
Ultrapurified water , Ion associate , Molybdosilicate , Flow Injection , Rhodamine B , Fluorescence quenching