Title of article :
Determination of microcystin-LR in surface water using high-performance liquid chromatography/tandem electrospray ionization mass detector
Author/Authors :
Zhang، نويسنده , , Lifeng and Ping، نويسنده , , Xiaofei and Yang، نويسنده , , Zhaoguang، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2004
Abstract :
The need for a rapid, sensitive, and reliable analytical method for microcystin-LR has been emphasized by the awareness of toxic cyanobacteria as a human-health risk through drinking water. Microcystin-LR is the most commonly reported microcystin which is produced by cyanobacteria. The WHO guideline for microcystin-LR in drinking water is 1 μg/l. In this paper, an effective method has been developed by application of high-performance liquid chromatography/tandem electrospray ionization mass detector in the determination of microcystin-LR in surface water sample. At the LC–MS–MS CID–full scan mode, different relative collision energies have been tested with 30% being used for further microcystin-LR analysis. The possible mass dissociation path has been proposed. Based on 30% relative collision energy, present method has an excellent method detection limit (MDL), which is as low as 2.6 ng/l. To the best of our knowledge, this represents one of the most sensitive methods in existence for the microcystin-LR analysis. This method has also been validated by evaluation of the calibration linearity, precision, accuracy, recovery, and mass ratio stability.
Keywords :
LC–MS–MS , electrospray ionization , microcystin-LR , Validation