• Title of article

    Voltammetric determination of azithromycin at the carbon paste electrode

  • Author/Authors

    Farghaly، نويسنده , , Othman Abd El-Moaty and Mohamed، نويسنده , , Niveen Abdel Latif and Kuroda، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2004
  • Pages
    8
  • From page
    531
  • To page
    538
  • Abstract
    Azithromycin (AZ) is the first member of a class of macrolide azalides antibiotics called azolides. A simple and selective square-wave voltammetric (SWV) method has been developed for the determination of azithromycin in pure form, in pharmaceutical preparation and in biological samples. Determination of azithromycin was accomplished with hand-make carbon paste electrode (CPE) in oxidative screen mode. The counter and reference electrodes were a Pt wire and a Ag/AgCl, respectively. Various parameters that can influence the peak signal (effect of buffer, ionic strength, accumulation time, pH and the composition of the paste) have been scrutinized. The best results were obtained in acetonitrile—aqueous 1 M sodium acetate–acetic acid buffer (pH 4.6) containing 0.1 M KCl (1:9; v/v) using a 15% paraffin oil CPE. The limits of detection and quantification of the pure drug are 0.463 and 1.544 ppb (with the correlation coefficient, r=0.9785and the standard deviation, S.D.=0.1 (n=5), for the accumulation time of 60 s), respectively. The method was successfully applied to the determination of the drug in urine and two forms of pharmaceutical formulations. Recoveries were 99.2—100.5% with S.D.=0.1—and 0.8% (n=5).
  • Keywords
    Dosage Forms , Carbon paste electrode , urine , Square-wave voltammetry of azithromycin
  • Journal title
    Talanta
  • Serial Year
    2004
  • Journal title
    Talanta
  • Record number

    1645409