• Title of article

    Application of Doehlert matrix to the study of flow injection procedure for selenium (IV) determination

  • Author/Authors

    Hellal، نويسنده , , F. and Dachraoui، نويسنده , , M.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2004
  • Pages
    6
  • From page
    1089
  • To page
    1094
  • Abstract
    A method for the rapid determination of selenium is reported. It is based on the injection of selenium sample into a stream of KI and subsequent measurement of the increase of the absorption at 290 nm due to the formation of I3−. A Doehlert matrix was applied for the determination of the optimal working conditions. With the developed method, a linear calibration curve from 1 to 10 ppm was obtained. For the analysis of semiconducting thin layers of CuInSe2 (CIS), a cationic resin was added to the sample solution in order to eliminate the effect of copper and indium on the absorbance values. With the developed procedure a sample throughput of 36 samples per hour and a relative standard deviation 1.1% was achieved.
  • Keywords
    FI , Selenium , Doehlert matrix , CUINSE2
  • Journal title
    Talanta
  • Serial Year
    2004
  • Journal title
    Talanta
  • Record number