Title of article :
Importance of REP values when comparing the CALUX bioassay results with chemoanalyses results: Example with spiked vegetable oils
Author/Authors :
Carbonnelle، نويسنده , , Sophie and Loco، نويسنده , , Joris Van and Overmeire، نويسنده , , Ilse Van and Windal، نويسنده , , Isabelle and Wouwe، نويسنده , , Nathalie Van and Leeuwen، نويسنده , , Stefan Van and Goeyens، نويسنده , , Leo، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2004
Pages :
5
From page :
1255
To page :
1259
Abstract :
Differences between chemical activated luciferase gene expression (CALUX) bioassay and chemoanalyses results are observed. aper shows that calculations of the TEQ values using REP values instead of WHO TEF values give different results. The REP values do affect the results obtained by the CALUX technique. These differences are more marked for the dioxin like PCB compounds (CALUX TEQ values are lower than WHO TEQ values) than for the dioxin compounds (CALUX TEQ values are higher than WHO TEQ values). LUX results were compared with the concentrations of the congeners’ spiked into the oil.
Keywords :
Chemical activated luciferase gene expression (CALUX) , Dioxins , PcBs , Relative potencies (REP)
Journal title :
Talanta
Serial Year :
2004
Journal title :
Talanta
Record number :
1646179
Link To Document :
بازگشت