Title of article :
Automated and manual luminescent assay of antioxidant capacity: analytical features by comparison
Author/Authors :
Girotti، نويسنده , , S. and Ferri، نويسنده , , E. and Fini، نويسنده , , F. and Bolelli، نويسنده , , L. and Sabatini، نويسنده , , A.G. and Budini، نويسنده , , R. and Sichertova، نويسنده , , D.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2004
Pages :
6
From page :
665
To page :
670
Abstract :
The analytical performances of a manual and a partially automated chemiluminescent (CL) assay, of total antioxidant capacity (TAC) were assessed. In both cases the light emitting reaction involved luminol, horseradish peroxidase and hydrogen peroxyde, but the emission kinetics and the parameters taken into account to calculate TAC values were completely different. The major characteristics expressing the quality of the two analytical methods, i.e. inaccuracy, repeteability and reproducibility, sensitivity, time required for the analysis and detection limit, were estimated by using standard solutions of Trolox. The reliability of the automated method, in comparison with the more validated manual one, was demonstrated testing food samples such as honey, wine and dietary supplements and performing a statistical analysis of the results. The comparison of the two series of data by t-test resulted in p values in the range 0.1–0.01. The time required for the analysis of each sample was reduced to one third using the automated method.
Keywords :
Microplate luminometer , antioxidant capacity , foods , Chemiluminescence
Journal title :
Talanta
Serial Year :
2004
Journal title :
Talanta
Record number :
1646561