Title of article :
Development of a sequential injection system for trace mercury determination by cold vapour atomic absorption spectrometry utilizing an integrated gas–liquid separator/reactor
Author/Authors :
Anthemidis، نويسنده , , Aristidis N. and Zachariadis، نويسنده , , George A. and Stratis، نويسنده , , John A.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2004
Pages :
5
From page :
1053
To page :
1057
Abstract :
A simple and robust time-based on-line sequential injection system for trace mercury determination via cold vapour atomic absorption spectrometry (CVAAS), employing a new integrated gas–liquid separator (GLS), which in parallel operates as reactor, was developed. Sample and reductant are sequentially loaded into the GLS while an argon flow delivers the released mercury vapour through the atomic absorption cell. The proposed method is characterized by the ability of successfully managing variable sample volume up to 30 ml in order to achieve high sensitivity. For 20 ml sample volume, the sampling frequency is 25 h−1. The calibration curve is linear over the concentration range 0.05–5.0 μg l−1 of Hg(II), the detection limit is cL = 0.02 μg l−1, and the relative standard deviation is sr = 2.6% at 1.0 μg l−1 Hg(II) level. The performance of the proposed method was evaluated by analyzing certified reference material and applied to the analysis of natural waters and biological samples.
Keywords :
mercury , Gas–liquid separator , Sequential injection , Cold vapour atomic absorption spectrometry
Journal title :
Talanta
Serial Year :
2004
Journal title :
Talanta
Record number :
1646754
Link To Document :
بازگشت