Title of article
Determination of Lead(II), Copper(II), Zinc(II), Cobalt(II), Cadmium(II), Iron(III), Mercury(II) using sequential injection extractions
Author/Authors
van Staden، نويسنده , , J.F. and Taljaard، نويسنده , , R.E.، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2004
Pages
10
From page
1203
To page
1212
Abstract
Sequential injection analysis is still dominated by single component analysis. In this proposed robust, economical simple instrumental system seven different metal ions are determined simultaneously using thin-film sequential injection extraction (SIE) with multivariate calibration and multiwavelength detection. Dithizone, in ethanol, is used as extractant and the metal dithizonates’ spectra are generated by a diode array spectrophotometer between 300 and 700 nm. The SI thin-film extraction using water miscible with ethanol works due to the hydrophobic interaction of ethanol with the Teflon wall to create a thin film. A sample frequency of 27 samples per hour was obtained with a sample carry-over of less than 1%. The results of the proposed sequential injection extraction system compare favourably with the results obtained by using standard atomic absorption spectrometry (AAS) methods on conventional extraction samples.
Keywords
SI thin-film extraction , Sequential injection extractions , Diode array spectrophotometer