• Title of article

    Chemiluminometric determination of carvedilol in a multi-pumping flow system

  • Author/Authors

    Pires، نويسنده , , Cherrine K. and Marques، نويسنده , , Karine L. and Santos، نويسنده , , Joمo L.M. and Lapa، نويسنده , , Rui A.S. and Lima، نويسنده , , José L.F.C. and Zagatto، نويسنده , , Elias A.G.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2005
  • Pages
    6
  • From page
    239
  • To page
    244
  • Abstract
    In this work a simple, fast, sensitive and selective flow-based procedure for the chemiluminometric determination of carvedilol, a recent non-cardioselective β-blocker with noteworthy antioxidant activity, is proposed. The developed methodology takes advantage of the antioxidant capacity of carvedilol to inhibit the chemiluminescence response resulting from the oxidation of luminol by hypochlorite, by acting as a hypochlorite scavenger. The analytical process was implemented in a multi-pumping flow system that employs multiple solenoid actuated micro-pumps as the only active components. These acted as solution insertion, propelling and commuting units assuring an easily controlled, low cost, compact and reliable analytical system. ar working range for carvedilol concentrations ranging from 1.2 × 10−7 to 3.0 × 10−6 mol l−1 (r > 0.999, n = 6), was obtained, with a detection limit of 8.7 × 10−9 mol l−1. The system handles about 65 samples per hour yielding precise results (R.S.D. < 1.3%, n = 10). Recoveries within 95 and 104% were obtained.
  • Keywords
    Multi-pumping flow system , Chemiluminescence , Carvedilol , Luminol
  • Journal title
    Talanta
  • Serial Year
    2005
  • Journal title
    Talanta
  • Record number