Title of article
Determination of substituted benzenes in water samples by fiber-in-tube liquid phase microextraction coupled with gas chromatography
Author/Authors
Wang، نويسنده , , Jun-Xia and Jiang، نويسنده , , Dong-Qing and Yan، نويسنده , , Xiu-Ping، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2006
Pages
6
From page
945
To page
950
Abstract
A method for determination of toluene, ethylbenzene, p-xylene, o-xylene, 1,3,5-trimethylbenzene and 1,2,4-trimethylbenzene in water samples was developed by a fiber-in-tube liquid phase microextraction technique (fiber-in-tube LPME) coupled with GC-flame ionization detector (FID). The method used a tube packed with polytetrafluoroethylene (PTFE) fibers as an extraction medium, improving the stableness of the solvent and the performance of extraction. Certain amounts of curled PTFE fibers were packed into a section of PTFE tube. Because the fibers were curled, they formed network structure in the tube. The fiber packed tube was firstly immersed into organic solvent to be filled with organic solvent and then was exposing to an aqueous solution to extract the target compounds. The extract was then retracted by a conventional GC microsyringe and analyzed by GC-FID. Extraction of the analytes in 8 ml aqueous solution for 15 min yielded enrichment factors of 224–361. The precision (R.S.D., n = 5) was 3.6–8.1% for peak area. The limit of detection (LOD, S/N = 3) for the six substituted benzenes were in the range of 0.3–5.0 μg l−1.
Keywords
Liquid phase microextraction , Substituted benzenes , Gas chromatography
Journal title
Talanta
Serial Year
2006
Journal title
Talanta
Record number
1649363