Title of article :
Preconcentration and speciation of chromium in drinking water samples by coupling of on-line sorption on activated carbon to ETAAS determination
Author/Authors :
Gil، نويسنده , , R.A. and Cerutti، نويسنده , , S. and Gلsquez، نويسنده , , J.A. and Olsina، نويسنده , , R.A and Martinez، نويسنده , , L.D.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2006
Abstract :
An on-line flow injection (FI) preconcentration-electrothermal atomic absorption spectrometry (ETAAS) method is developed for trace determination of chromium in drinking water samples by sorption on a conical minicolumn packed with activated carbon (AC) at pH 5.0. The chromium was removed from the minicolumn with 1.0% (v/v) nitric acid. An enrichment factor (EF) of 35-fold for a sample volume of 10 ml was obtained. The detection limit (DL) value for the preconcentration method proposed was 3.0 ng l−1. The precision for 10 replicate determinations at the 0.5 μg l−1 Cr level was 4.0% relative standard deviation (R.S.D.), calculate with the peak heights obtained. The calibration graph using the preconcentration system for chromium was linear with a correlation coefficient of 0.9992 at levels near the detection limits up to at least 50 μg l−1. The method was successfully applied to the determination of Cr(III) and Cr(VI) in drinking water samples.
Keywords :
Chromium speciation , Activated carbon , On-line preconcentration , ETAAS , Drinking water