Title of article
Molecularly imprinted TiO2 thin film using stable ground-state complex as template as applied to selective electrochemical determination of mercury
Author/Authors
Liu، نويسنده , , Zhihang and Huan، نويسنده , , Shuangyan and Jiang، نويسنده , , Jianhui and Shen، نويسنده , , Guoli and Yu، نويسنده , , Ruqin، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2006
Pages
6
From page
1120
To page
1125
Abstract
Molecular recognition sites for mercury ions were imprinted in TiO2 film using stable ground-state complex of 1-amino-8-naphtol-3,6-disodium sulfonate (ANDS) and mercury ions as template. The complex ratio between mercury ions and ANDS was estimated to be 2:1. Compared with the controlled and pure TiO2 electrodes, the imprinted electrode revealed selectivity towards the imprinted ions. Linear calibration plots for mercury ions were obtained and the regression equation was Ip (μA) = 4.29 × 10−7 + 19.40 [Hg2+] with a detection limit of 3.06 × 10−9 mol/l. The imprinted electrode could be used for more than 1 month. Recoveries were calculated at both high and low concentrations, with a mean recovery of 99.6%.
Keywords
Imprinted thin film , Stable ground-state complex , LSV , Sol–gel , Sensor
Journal title
Talanta
Serial Year
2006
Journal title
Talanta
Record number
1649433
Link To Document